| Model Number |
0210 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Reportable as this delayed patient treatment.
|
| |
|
Event or Problem Description
|
|
Nebulizer failures (medication bubbles but not aerosol and tube popping off).
|
| |
|
Event or Problem Description
|
|
Nebulizer failures (medication bubbles but not aerosol and tube popping off).
|
| |
|
Additional Manufacturer Narrative
|
|
Reportable as this delayed patient treatment.Complaint not confirmed due to lack of a photo or return.Reviewed the dhr and no anomalies were found.Received samples and all nebulized as expected.One of the samples had an unrelated fault that instigated ncmr-02911.Reviewed the complaint history and there was only one previous complaint for this part.There is no trending.Performed risk analysis with rma-20013b and the related risk is 6/10, less than the capa initiation threshold of 8.Sent a resolution to the customer.
|
| |
|