• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES EB240, VOYANT OPEN FUSION, 6 B/X; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES EB240, VOYANT OPEN FUSION, 6 B/X; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EB240
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit will not be returning to applied medical for evaluation.A follow-up report will be submitted upon completion of investigation.
 
Event or Problem Description
Procedure performed: low anterior resection (lar).Event description: the rep was present for the case, and it was an open case used with an alexis.On the very first activation the device did not seal.The jaws clamped both sides with ima in between.The generator beeped a number of times during the seal cycle and the end tones were heard but it took longer than usual to complete the seal cycle.The device did not seal at all.The vessel was cut and tied with sutures.The device was not used for some time as the surgeon was teaching how to cut & tie with sutures.Later in the case, the device was used again and for the second and third activations it worked fine.They completed the case with the same device.There was no adverse patient injury.The product will not return.Additional information was received via email on 10nov2022 from [voyant project manager, applied medical]: "no alarms sounded".The blade was activated after hearing the end tone.Yes, bleeding was experienced after activating the blade.The patient did not exhibit any vascular pathology, nor was the device activated on diseased or inflamed tissue.Zero tissue effects were observed at the seal site.Type of intervention: cut & tied with sutures.Patient status: no adverse patient injury.
 
Additional Manufacturer Narrative
The event unit was not returned to applied medical for evaluation.As the event unit was not returned, applied medical is unable to determine if the event unit exhibited any non-conformances that could have contributed to the reported event.In the absence of the event unit, it is difficult to determine if the reported event was caused by a manufacturing non-conformance or circumstantial factors at the time of use.Applied medical has reviewed the details surrounding the event and is unable to determine the exact root cause of the event.
 
Event or Problem Description
Procedure performed: low anterior resection (lar).Event description: the rep was present for the case, and it was an open case used with an alexis.On the very first activation the device did not seal.The jaws clamped both sides with ima in between.The generator beeped a number of times during the seal cycle and the end tones were heard but it took longer than usual to complete the seal cycle.The device did not seal at all.The vessel was cut and tied with sutures.The device was not used for some time as the surgeon was teaching how to cut & tie with sutures.Later in the case, the device was used again and for the second and third activations it worked fine.They completed the case with the same device.There was no adverse patient injury.The product will not return.Additional information was received via email on 10nov2022 from [voyant project manager, applied medical]: "no alarms sounded".The blade was activated after hearing the end tone.Yes, bleeding was experienced after activating the blade.The patient did not exhibit any vascular pathology, nor was the device activated on diseased or inflamed tissue.Zero tissue effects were observed at the seal site.Type of intervention: cut & tied with sutures.Patient status: no adverse patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EB240, VOYANT OPEN FUSION, 6 B/X
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15830541
Report Number2027111-2022-00828
Device Sequence Number17371845
Product Code GEI
UDI-Device Identifier00607915138455
UDI-Public(01)00607915138455(17)241004(30)01(10)1463008
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K201212
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2024
Device Model NumberEB240
Device Catalogue Number101474615
Device Lot Number1463008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date11/19/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALEXIS
Patient SexUnknown
-
-