Model Number 9-ASD-022
Medical Device Problem Codes Difficult to Fold, Unfold or Collapse (1254); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, a 22mm amplatzer septal occluder was selected for an implant using a 10 f amplatzer trevisio intravascular delivery system.During the procedure, the device was presented in a bulbous deformation.There was no interaction with cardiac structures during deployment.There was no angulation or kink noticed in the delivery system.It was note that the 22mm amplatzer septal occluder was ultimately implanted.The patient status was noted as stable.There was no additional information was provided.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of device deformity was reported.A returned device inspection could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported incident could not be conclusively determined, but may have been caused by attempting to implant the device with a larger than recommended delivery system.There is no indication for a product quality issue with regards to manufacture, design, or labeling.Please note, per the amplatzer septal occluder instructions for use, that a 9 french sheath is recommend for use with a 22mm septal occluder.A 10 french sheath was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15830848
Report Number2135147-2022-02104
Device Sequence Number12031684
Product Code MLV
UDI-Device Identifier05415067019295
UDI-Public05415067019295
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model Number9-ASD-022
Device Catalogue Number9-ASD-022
Device Lot Number8090965
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date11/19/2022
Supplement Report FDA Received Date01/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexMale
-
-