• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TRIFOCAL IOL; LENS, MULTIFOCAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ PANOPTIX TRIFOCAL IOL; LENS, MULTIFOCAL Back to Search Results
Model Number TFNT00
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported that during implantation of an intraocular lens (iol) the first lens was injected with no difficulty into the eye, noted a linear crack near the centre of the optic and removed it.Opened the spare lens, had a quick look under the microscope but didn't see any crack in the lens (i have to say that i didn't view it under high magnification).Loading and injection of lens was easy with no issue.Once again, saw the same linear crack near the centre of the optic.Didn't have any more lens so i left it in the eye following a discussion with the patient.There was no resistance with the latter and lens went in smoothly.I have used the injector many times in the past with no issue.Additional information was requested and received sating there was no patient harm.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Photo was provided.Photo #2 was indicated to be the implanted lens.The photo showed a multifocal single-piece lens in the eye.What appeared to be a crack was observed off center of the optic.The crack was perpendicular to the fold path.This is similar in appearance to damage that may occur if there is a plunger over/under ride of the optic during delivery.A qualified cartridge and handpiece were indicated with a non-qualified viscoelastic.The root cause for the reported crack could not be determined.The lens remains implanted.What appeared to be a crack was observed in the provided photo.The area was similar in appearance to damage that may occur if there is a plunger over/under ride of the optic during delivery.A determination could not be made from the photo.A non-qualified viscoelastic was indicated.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu(instruction for use) instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15830896
Report Number1119421-2022-02443
Device Sequence Number5103615
Product Code MFK
UDI-Device Identifier00380652357791
UDI-Public00380652357791
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTFNT00
Device Catalogue NumberTFNT00.230
Device Lot Number15382977
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer11/15/2022
Initial Report FDA Received Date11/19/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/07/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL CARTRIDGE D; MONARCH IOL , INJECTOR; VISTHESIA
Patient Age45 YR
Patient SexFemale
-
-