ZIMMER BIOMET SPINE INC. CORD 300 MM; THE TETHER - VERTEBRAL BODY TETHERING SYSTEM Back to Search Results
Catalog Number 204H0300
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed to replace a tether cord that broke in two places post-operatively after the patient went horseback riding.
 
Additional Manufacturer Narrative
Procode: qhp.Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Additional Manufacturer Narrative
Corrections in d4: udi number and h3.Additional information in d4: expiration date, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection revealed that the cord was fractured in multiple pieces.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device use: this device is used for treatment.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Event or Problem Description
It was reported that a revision surgery was performed to replace a tether cord that broke in two places post-operatively after the patient went horseback riding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORD 300 MM
Common Device Name
THE TETHER - VERTEBRAL BODY TETHERING SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key15831053
Report Number3012447612-2022-00268
Device Sequence Number14976562
Product Code QHP
UDI-Device Identifier00880304865068
UDI-Public(01)00880304865068(17)230831(10)2966796
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
H190005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Catalogue Number204H0300
Device Lot Number2966796
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date11/19/2022
Supplement Report FDA Received Date02/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;