| Catalog Number |
204H0300 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed to replace a tether cord that broke in two places post-operatively after the patient went horseback riding.
|
| |
|
Additional Manufacturer Narrative
|
|
Procode: qhp.Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d4: udi number and h3.Additional information in d4: expiration date, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation: visual inspection revealed that the cord was fractured in multiple pieces.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device use: this device is used for treatment.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.A follow-up report will be submitted if new information is received that changes the information provided in this report.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed to replace a tether cord that broke in two places post-operatively after the patient went horseback riding.
|
| |
|