• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. 4K UHD LCD MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. 4K UHD LCD MONITOR Back to Search Results
Model Number OEV321UH
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/03/2022
Event Type  malfunction  
Manufacturer Narrative
Olympus has not yet received the device from the customer for evaluation.The customer did report that the screen continued not to work after the cable was changed.This investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer contacted olympus to report a failure to power on with their 4k uhd lcd monitor.The failure to power on was discovered before a procedure, and no patient or user harm has been reported.
 
Event Description
The intended procedure has been reported to have been diagnostic.
 
Manufacturer Narrative
The subject device has been returned to an olympus repair center for evaluation and inspection, and the customer's reported problem has been confirmed; the power supply was found to be faulty.This report is also being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than a year since the subject device was manufactured.Based on the results of the investigation, it is likely that the faulty power supply led to the malfunction.A definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K UHD LCD MONITOR
Type of Device
MONITOR
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15831095
MDR Text Key307614208
Report Number3002808148-2022-04418
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170418723
UDI-Public04953170418723
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K220069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOEV321UH
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-