• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES Back to Search Results
Model Number MAJ-2330
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus technical assistance center (tac), there were multiples oer-elite connectors that were ¿bad¿.The customer reported that it was an out of box failure.There was no patient involvement.
 
Additional Manufacturer Narrative
Related patient identifiers: the customer reported that multiple connecting tubes were bad.(b)(6) (addresses model number maj-2110; lot# 0ya), (b)(6) (addresses model number maj-2112; lot# 0za), (b)(6) (addresses model number maj-2113; lot# 0ya) (b)(6) (addresses model number maj-2114; lot# 0ya, (b)(6) (addresses model number maj-2115; lot# 0xa and (b)(6) (addresses model number maj-2330; lot# 0xa) the customer reported that the connectors will be returned.However, to date, the connecting tube has not been returned.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Additional Manufacturer Narrative
This report is being submitted to provide additional information based on the legal manufacturer¿s final investigation.Since the manufacturing date of the device is unknown, the device history record was unable to be reviewed, however, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the device investigation, the cause of the reported ¿bad¿ oer-elite connectors issue was confirmed.The root cause of the damaged connector could not be determined, however, the issue was likely the result of an external load applied to the cleaning tube, making it impossible to connect.An external load on the irrigation tube may be caused by the user applying force in the wrong direction.The event can be prevented by following the instructions for use which state: ¿abnormalities can be detected by conducting the following inspections: 9 preparation and inspection 2 move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken¿.Olympus will continue to monitor field performance for this device.
 
Additional Manufacturer Narrative
This report is being submitted to provide additional information from customer follow-up, g3 correction and device manufacturing date.B5: updated with information received from the customer.Correction to g3 of the initial medwatch.The aware date should be 29-aug-2022.The subject device was manufactured in october, 2010 but an exact date is unavailable as the complete serial number was not provided.
 
Event or Problem Description
Customer confirmed all the connectors broke at the metal clips.
 
Additional Manufacturer Narrative
This report is being supplemented to provide a correction to the legal manufacturer's final investigation.Based on the results of the corrected investigation, there has since been a design change to prevent reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Common Device Name
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15831102
Report Number9610595-2022-04321
Device Sequence Number17404379
Product Code FEB
UDI-Device Identifier04953170404207
UDI-Public04953170404207
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K190969
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-2330
Device Lot Number0XA
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2022
Supplement Date Received by Manufacturer01/05/2023
02/03/2023
03/31/2023
Initial Report FDA Received Date11/19/2022
Supplement Report FDA Received Date01/25/2023
02/09/2023
04/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-