Model Number OCCLUDER
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, a 5mm x 6mm amplatzer piccolo occluder was chosen for implant.The sizing of the device was determined via echocardiography and fluoroscopy imaging.Roughly 7 minutes after release, the occluder embolized into the pulmonary artery.The implanter believes the pliable duct didn't allow for an accurate measurement of the defect.The measured dimensions of the defect are: 8mm length x 3.4mm diameter.There was an attempt to snare the device but it was unsuccessful and the patient underwent surgery to remove the device.This was considered an emergency procedure.It was noted that the patient's status was noted as stable and the surgery was successful.The patient remained hemodynamically stable throughout the procedure.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Additional Manufacturer Narrative
An event of device embolization was reported.A returned device assessment could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Subsequent to the previously filed report, additional information was received that the patient was at least 3 days old and was over 2kg at the time of the procedure.The device was placed extraductal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PICCOLO
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15831476
Report Number2135147-2022-02115
Device Sequence Number5298710
Product Code MAE
UDI-Device Identifier05415067031051
UDI-Public05415067031051
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOCCLUDER
Device Catalogue Number9-PDAP-05-06-L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/22/2022
Supplement Date Received by Manufacturer01/16/2023
Initial Report FDA Received Date11/20/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ABBOTT DELIVERY SYSTEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 DA
Patient SexUnknown
Patient Weight2 KG
-
-