• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 19 CM RETR; CATHETER, HEMODIALYSIS, IMPLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-L 15 FR X 19 CM RETR; CATHETER, HEMODIALYSIS, IMPLA Back to Search Results
Model Number IPN001348
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
The complaint is reported as: we observed a rigidity more important than usual on the "tubing" of the catheter which leads to a kink of the catheter linked to the clamps.On (b)(6) 2022, a patient faced this issue, with dialyses machine that stopped several times beacause of negative blood pressure (-300mmhg).This affects the blood flow, the quality of dialysis, the increase in the risk of coagulation, the patient's discomfort and delay in the department.
 
Event or Problem Description
The complaint is reported as: we observed a rigidity more important than usual on the "tubing" of the catheter which leads to a kink of the catheter linked to the clamps.On (b)(6) 2022, a patient faced this issue, with dialyses machine that stopped several times beacause of negative blood pressure (-300mmhg).This affects the blood flow, the quality of dialysis, the increase in the risk of coagulation, the patient's discomfort and delay in the department.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-L 15 FR X 19 CM RETR
Common Device Name
CATHETER, HEMODIALYSIS, IMPLA
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15831585
Report Number9680794-2022-00763
Device Sequence Number14733383
Product Code MSD
UDI-Device Identifier10801902127171
UDI-Public10801902127171
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141051
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberIPN001348
Device Catalogue NumberCS-15192-VFE
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer12/14/2022
Initial Report FDA Received Date11/20/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEMODIALYSIS.; HEMODIALYSIS.
Patient SexUnknown
-
-