H10: as this malfunction is considered one event identified by the dealer; only one mdr report will be submitted for the reported quantity affected of thirty for this event.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: three bard ptfe pledgets package were returned for evaluation.During visual evaluation, no evidence of breach of sterile barrier was observed.In sample one, yellowish particle was observed under a microscope, while a black particle was observed in sample two.Black and yellowish particles were found in sample three.Therefore, the investigation is confirmed for the reported contamination issue.A definitive root cause for the contamination issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 12/2026), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|