ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 40MM; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Migration or Expulsion of Device (1395); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: mdr328079, mdr328081, mdr328085.Concomitant medical products: item number 51-106180 item name tprlc 133 mp type1 pps so 18.0 lot number: 6185849, item number 010000749 item name g7 neutral arcomxl lnr 40mm h lot number 6139276, item number s331140 item name selex/magnum mod hd 40mm+3 lot number 724250.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, as supplemental medwatch will be submitted.
 
Event or Problem Description
It was reported that 10 months after a left total arthroplasty, the patient was revised due to groin pain, loosening and migration of the acetabular cup.The head, liner, shell and screws were all placed without complication.The stem was stable and remained intact.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4; g3; h2; h3; h4; h6.Reported event was confirmed via medical records provided and reviewed by a health care professional.A review of the device history records identified no related deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information on the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 40MM
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15832945
Report Number0001825034-2022-02630
Device Sequence Number12270685
Product Code LZO
UDI-Device Identifier00880304527492
UDI-Public(01)00880304527492(17)270206(10)3970990
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number010001001
Device Lot Number3970990
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer01/27/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE H10 FOR ALL CONCOMITANT DEVICES
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
Patient Weight114 KG
-
-