• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 FREEDOM CONST E1 LNR 36MM G; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 FREEDOM CONST E1 LNR 36MM G; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the liner would not seat in the osseoti shell.A new shell was required to complete the procedure.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: mdr328146 and mdr328148.Item number: 00625006525, item name: bone scr 6.5x25 self-tap, lot #: j7222341; item number: 31-323230, item name: 3.2mmx30mm rnglc+ acet drl bit, lot #: 731380; item number: 00625006520, item name: bone scr 6.5x20 self-tap, lot #: j7278081; item number: 00625006530, item name: bone scr 6.5x30 self-tap, lot #: j7007847; item number: 00625006540, item name: bone scr 6.5x40 self-tap, lot #: j7332696; item number: 110010269, item name: g7 osseoti multihole 62mm h, lot #: 6902757; item number: 802403606, item name: zb 12/14 cocr frdm 36mm x +9, lot #: 3062343.Item number 00625006515 item name bone scr 6.5x15 self-tap lot # j6867004 item number 00625006515 item name bone scr 6.5x15 self-tap lot # j6993494 item number 00625006530 item name bone scr 6.5x30 self-tap lot # j6960420 item number 30253608 item name g7 vit e frdm const +5mm 36h lot # 65330653 the product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
There is no update to the previous event description provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified both of the liners had damage to the locking feature.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: the image provided was of the final implanted products.They did not enhance the investigation as the reported issue was to do with the returned components.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 FREEDOM CONST E1 LNR 36MM G
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15833021
Report Number0001825034-2022-02636
Device Sequence Number17499556
Product Code PBI
UDI-Device Identifier00880304989092
UDI-Public(01)00880304989092(17)260304(10)6982165
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2026
Device Model NumberN/A
Device Catalogue Number010000985
Device Lot Number6982165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer02/14/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date02/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexFemale
-
-