ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL VASCULAR LIN STAP; STAPLE, IMPLANTABLE Back to Search Results
Model Number TX30V
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure, removed from the surgery field because the procedure was not opened during the test before it was used.Fired to open the approximated jaw outside the surgical field.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/08/2022.Additional information received: during the test before use, it did not open after approximation, so it was removed from the surgical field without use.Fired while opening the approximated jaw outside the surgical field.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 01/13/2023.D4: batch # 257a60.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the tx30v device arrived with no apparent damaged and with a reload loaded on the device.In addition, a tyvek was received along with the device.The reload was received with only 9 staples present and protruding.Also, the drivers were noted partially advanced.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria.It should be noted that when manipulating the device only the closing trigger should be grasped until ready to fire the device.Do not grasp the firing trigger before the device is to be fired.If the firing trigger is manipulated it could cause the staples to deploy partially or completely resulting in malformed staples and a premature lockout situation.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch 257a60, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15833127
Report Number3005075853-2022-08084
Device Sequence Number14976573
Product Code GDW
UDI-Device Identifier10705036003335
UDI-Public10705036003335
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberTX30V
Device Catalogue NumberTX30V
Device Lot Number359A46
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/23/2022
Supplement Date Received by Manufacturer11/23/2022
12/26/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/09/2022
01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-