| Model Number |
I400+ |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable iron2 iron gen.2 results for 2 patient samples on a cobas integra 400 plus module.On (b)(6) 2022, patient 1 had an initial iron result of 5.82 ug/dl.The first repeat result was 7.84 ug/dl and the second repeat result was 42.98 ug/dl.The sample was sent to another lab and the iron result was 13.19 ug/dl.On (b)(6) 2022, patient 2 had an initial iron result of 61.74 ug/dl.The first repeat result was 0.31 ug/dl.On (b)(6) 2022, the sample was repeated a second time and the result was 12.83 ug/dl.No questionable results were reported outside of the laboratory.The iron reagent lot is 65786901 and the expiration date is 31-aug-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Iron calibration performed on 29-sep-2022 was acceptable.Calibration was not performed again until 18-oct-2022.The investigation found that there were several postponed calibration messages in the data log.The message log showed that controls performed on 17-oct-2022 were acceptable but calculated using a postponed calibration.The alarm trace showed several clot detection and no fluid detection alarms.The investigation did not identify a product problem.The root cause of the event could not be determined.
|
| |
|