MAUDE Adverse Event Report: INTRINSIC THERAPEUTICS, INC BARRICAID ACD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTRINSIC THERAPEUTICS, INC BARRICAID ACD Back to Search Results
Model Number BAR-A8-10MM
Medical Device Problem Codes Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
Surgeon inserting the barricaid implant and noticed the mesh starting to buckle.The surgeon then removed the failed implant with mesh buckling and attempted a second implantation using the same damaged implant.The buckled mesh then contineud to buckle and it detached from the anchor.Then the surgeon removed the now separated mesh and sunk the implant further into the vertebral body.Anchor remains implanted and the patient was closed up.
 
Additional Manufacturer Narrative
Surgeon was interviewed after the surgery, and they agreed that they were using an angle that was too steep for the initial insertion.The surgeon was instructed when the buckling was noticed, to replace the damaged implant and was offered a new implant by the intrinsic representative present at the time of the operation.They decided to continue with using the damaged implant which caused it to fail by having the mesh barrier (seal) to detach from the anchor.This failure of the device was due to improper technique initially and continuing to use a damaged implant.There was no patient harm associated with this adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARRICAID ACD
Common Device Name
BARRICAID
Manufacturer (Section D)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer (Section G)
INTRINSIC THERAPEUTICS, INC
30 commerce way
woburn MA 01801
Manufacturer Contact
rachael nelsen
30 commerce way
woburn, MA 01801
7819320222
MDR Report Key15833355
Report Number3006232063-2022-00022
Device Sequence Number5391666
Product Code QES
UDI-Device IdentifierM906BARA810MM0
UDI-PublicM906BARA810MM0
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P160050
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/21/2024
Device Model NumberBAR-A8-10MM
Device Catalogue Number2731015-A8
Device Lot Number04302120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/17/2022
Initial Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexUnknown
-
-