• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION THREADER; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION THREADER; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 1860
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 80% stenosed target lesion was located in the moderately tortuous and mildly calcified first obtuse marginal 1.A 1.2mm x 12mm threader balloon catheter was advanced for dilatation.However, during the first inflation at 14 atmospheres for 15 seconds, the balloon ruptured.The device was removed without any problem and the procedure was completed successfully with another 1.2x12 threader balloon catheter.There were no patient complications nor injuries reported.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 80% stenosed target lesion was located in the moderately tortuous and mildly calcified first obtuse marginal 1.A 1.2mm x 12mm threader balloon catheter was advanced for dilatation.However, during the first inflation at 14 atmospheres for 15 seconds, the balloon ruptured.The device was removed without any problem and the procedure was completed successfully with another 1.2x12 threader balloon catheter.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluation by mfr: the returned product consisted of a threader mr balloon catheter.The device was visually and microscopically examined.There was contrast and blood present in the inflation lumen and balloon.There was blood in the guidewire lumen, and the balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole in the balloon located at the distal end of the markerband.The device failed to inflate to rated burst pressure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREADER
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15833461
Report Number2124215-2022-47936
Device Sequence Number14961120
Product Code LOX
UDI-Device Identifier08714729848127
UDI-Public08714729848127
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K134031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/11/2024
Device Model Number1860
Device Catalogue Number1860
Device Lot Number0030132204
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer12/14/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-