• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D010
Medical Device Problem Codes Failure to Convert Rhythm (1540); Inaccurate Synchronization (1609)
Health Effect - Clinical Code Ventricular Fibrillation (2130)
Date of Event 11/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this patient experienced ventricular tachycardia (vt).This implantable cardioverter defibrillator (icd) delivered anti-tachycardia pacing (atp) and one 41 joule shock.After the shock, the patient's rhythm accelerated to ventricular fibrillation (vf).The next seven shocks failed to convert the rhythm and therapy was exhausted.This patient was hospitalized and required external defibrillation.Technical services (ts) analyzed device episode data and recommended a dual coil tachy lead.It was noted that defibrillation threshold (dft) test was performed successfully.No additional adverse patient effects were reported.Currently, this icd remains in service.
 
Event or Problem Description
It was reported that this patient experienced ventricular tachycardia (vt).This implantable cardioverter defibrillator (icd) delivered anti-tachycardia pacing (atp) and one 41 joule shock.After the shock, the patient's rhythm accelerated to ventricular fibrillation (vf).The next seven shocks failed to convert the rhythm and therapy was exhausted.This patient was hospitalized and required external defibrillation.Technical services (ts) analyzed device episode data and recommended a dual coil tachy lead.It was noted that defibrillation threshold (dft) test was performed successfully.No additional adverse patient effects were reported.Currently, this icd remains in service.
 
Additional Manufacturer Narrative
After medical review it was determined to add code for life-threatening event that required external intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN MINI ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112 5798
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15833509
Report Number2124215-2022-48115
Device Sequence Number12110139
Product Code LWS
UDI-Device Identifier00802526533785
UDI-Public00802526533785
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/24/2017
Device Model NumberD010
Device Catalogue NumberD010
Device Lot Number206671
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/04/2022
Supplement Date Received by Manufacturer12/14/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/24/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age74 YR
Patient SexMale
-
-