| Model Number |
NOV 6.5MM HEX DRIVER ARTICULATING |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported during surgical use, the surgeon performed tha procedure and during this surgery, surgeon encountered a incident that a tip of driver was disassembled and all parts except for one shaft was retrieved from surgical field and operation room.Using the imaging machine, he checked if a shaft was left in patient body.As a result, there was not found out a shaft in patient body.There was a 31-45 min surgical delay/prolongation.There was no adverse event as a result of the surgical delay.Patient was last known to be in stable condition following the event.Image received.Sales rep was unable to obtain x-ray.Device is available for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g6, h2, h3 and h6 have been updated accordingly.(h3) the broken tip driver reported may have been due to the press-fit dowel pin backing out as a result of numerous surgical uses and subsequent sterilization cycles, which led to ultimate disassembly of the tip and washer subcomponents from the device.The most probable root cause associated with the reported event of ¿disassembled / misassembled¿ is associated with the unwanted disassembly of the device, or incorrect assembly of the device (attaching mating instrumentation incorrectly, assembling an implant construct off-axis, etc.).
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: h11.Upon reassessment of the reported event, it was determined that the event did not involve a death or serious injury nor would be likely to cause or contribute to a death or serious injury if the reported malfunction were to recur.As a result, this complaint event is no longer considered reportable by exactech and this report may be disregarded.
|
| |
|