Model Number 08P1934
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer observed a falsely elevated magnesium result generated on an alinity c processing module for one patient.The following results were provided: sid (b)(6): initial: 0.43, repeated: 0.63, repeat on different analyzer: 0.40.There was no impact to patient management reported.
 
Additional Manufacturer Narrative
Similar incident information will be provided in the final report.
 
Additional Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.A search for similar complaints and ticket trending review did not identify any trends or issues with the complaint lot number or current complaint issue.Device history record review did not identify any non-conformances or deviations related to the complaint lot and complaint issue.Labeling was reviewed and was found to adequately address the issue under review.Based on the investigation, no systemic issue or deficiency was identified for the alinity c magnesium reagent, lot 40152ud00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key15833618
Report Number3005094123-2022-00266
Device Sequence Number5391676
Product Code JGJ
UDI-Device Identifier00380740176532
UDI-Public00380740176532
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/22/2023
Device Model Number08P1934
Device Catalogue Number08P19-34
Device Lot Number40152UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/22/2022
Supplement Date Received by Manufacturer01/13/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/05/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, AC01907; ALNTY C PROCESSING MODU, 03R67-01, AC01907
Patient SexUnknown
-
-