• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. LOW PROFILE SCREW, 1.4X8MM, CORTICAL, TI; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. LOW PROFILE SCREW, 1.4X8MM, CORTICAL, TI; PLATE, FIXATION, BONE Back to Search Results
Model Number LOW PROFILE SCREW, 1.4X8MM, CORTICAL, TI
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Event or Problem Description
On 10/31/2022, it was reported by a sales representative via phone that a ar-18714-08, ar-18714-07, and a ar-18714-09 screw broke.This occurred during a phalanx fracture on (b)(6) 2022 when inserting the ar-18714-08 screw it broke about 3/4 of the way, the surgeon then attempted to insert a ar-18714-09 screw to secure the plate to the bone, upon insertion the screw broke.Then a ar-18714-07 screw was used to secure the plate as it was already down, when attempting to secure the plate it broke.All broken fragments remain in the patient.The patient had very good bone quality and the sockets were prepped for insertion with a ar-18700-06 drill bit.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOW PROFILE SCREW, 1.4X8MM, CORTICAL, TI
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15833639
Report Number1220246-2022-05768
Device Sequence Number14976577
Product Code HRS
UDI-Device Identifier00888867303874
UDI-Public00888867303874
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K191326
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLOW PROFILE SCREW, 1.4X8MM, CORTICAL, TI
Device Catalogue NumberAR-18714-08
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Initial Report FDA Received Date11/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-