Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Pictures were sent for review.From the pictures provided, the investigator was unable to be confirm the failure mode reported as "inflation issue - unable to inflate".Part number reported is not being manufactured currently, however, another part number from the same family was use for the "verification of failure mode reported in the current manufacturing process" and was conducted as follows: 80 samples were taken from the current production, the cuff from the samples were inflated and left for four hours, after this time samples were checked to verify if any leak was present however all samples were still fully inflated, defect reported "inflation issue - unable to inflate" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the availa ble information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
|
It was reported "once the tube entered the patient, the balloon did not inflate, apparently it had some leak." patient condition unknown at time of report.Additional information has been requested to complainant, no response at time of report.
|