| Model Number |
00-2490-032-00 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Foreign: south africa.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the threaded tip of the instrument fractured when the surgeon attempted to implant the implant into the patient.No pieces fell into the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6.H6: component codes: mechanical(g04)- impactor.Visual examination of the provided picture identified the threaded tip of the impaction head has fractured and was then retained in the targeting handle.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|