Model Number 1570-11-150
Medical Device Problem Codes Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cyst(s) (1800); Fall (1848); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Pain (1994); Discomfort (2330); Metal Related Pathology (4530); Swelling/ Edema (4577)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial reporter occupation: (b)(6).Component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the recovering following removal of depuy pinnacle hip due to complications including pseudotumor.Doi: unknown.Dor: unknown.Affected side: unknown.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Medical records were received and stated that on (b)(6) 2022, patient had a revision for right total hip replacement due to metallosis and localized fluid collection.Operative reported a darkish staining of metallic wear debris noted in anterior portion of acetabular device.
 
Event or Problem Description
Medical records received.Patient was revised due to the following pathologic findings: cyst, corrosion; metallosis.There was metallic staining at the base of the trunnion.Doi:(b)(6) 2009 - dor: (b)(6) 2022 (right hip).
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b1 (product problem), b5, d6a, h6 health effect - clinical code if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 medical device problem code.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary - no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot - a manufacturing record evaluation was performed for the finished device product code:157011150 lot number:d2thf1, and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ8 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15834000
Report Number1818910-2022-23353
Device Sequence Number6035444
Product Code LPH
UDI-Device Identifier10603295060093
UDI-Public10603295060093
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1570-11-150
Device Catalogue Number157011150
Device Lot NumberD2THF1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2022
Supplement Date Received by Manufacturer03/01/2023
08/17/2023
08/28/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date03/07/2023
08/24/2023
08/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP; ASPHERE M SPEC 12/14 44 +1.5; PINNACLE 300 ACET CUP 62MM; PINNACLE MTL INS NEUT44IDX62OD; SUMMIT POR TAPER SZ8 HI OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale