Model Number IPN924969
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reported issue: it was reported that the end of the tuohy has folded up on itself and the residents or attendings have been unable to pass the spinal needle or catheter through.Upon further questioning the doctor stated that she was not sure that anything broke/chipped off in the patient and that the resident that was placing the epidural claimed thought it chipped but i did not see anything at the end of the needle.The tip may have folded over itself.The bevel of needles appears flimsy , and i think they fold over themselves and occlude.She opened a second kit and placed a new epidural.The patient's current condition is unknown.
 
Additional Manufacturer Narrative
Qn# (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Additional Manufacturer Narrative
(b)(4).A device history record review could not be performed as no lot number was provided by the customer.A review of sales history data was performed to obtain a lot number.However, there was no record of purchase of this product by this customer.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review could not be performed as no lot number was provided by the customer.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without the sample.No further action is required at this time.
 
Event or Problem Description
Reported issue: it was reported that the end of the tuohy has folded up on itself and the residents or attendings have been unable to pass the spinal needle or catheter through.Upon further questioning the doctor stated that she was not sure that anything broke/chipped off in the patient and that the resident that was placing the epidural claimed thought it chipped but i did not see anything at the end of the needle.The tip may have folded over itself.The bevel of needles appears flimsy , and i think they fold over themselves and occlude.She opened a second kit and placed a new epidural.The patient's current condition is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPINAL ANES/EPIDURAL CATH KIT
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
effie jefferson
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key15834300
Report Number1036844-2022-00069
Device Sequence Number20190376
Product Code BSO
UDI-Device Identifier10801902209556
UDI-Public10801902209556
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN924969
Device Catalogue NumberAK-05560
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2022
Supplement Date Received by Manufacturer01/04/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED; NONE REPORTED