EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
LOGIC CR TIB INSERT SLOPE+, SZ 3, 9MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/31/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, approximately 6.5 years post initial tka, the female patient femur had an increased posterior dwell point- whether the sloped poly or a compromised pcl was the root cause is unclear.This likely led to the early delamination of the posterolateral aspect of the poly bearing.There was also significant osteolysis under the posterior condyles- although the femur wasn¿t grossly loose.Patient was revised to a competitor's devices.Per sales rep "we have seen a similar trend in our complaints/retrieval database of aged/sloped cr poly bearings.".
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: logic cr femoral cem, right, sz 3 (cat# 02-010-03-0330 / serial# (b)(4).Lgc tibial fit tray cem sz 3f / 3t (cat# 02-012-45-3030 / serial# (b)(4).Three peg patella 35mm (cat# 200-02-35 / serial# (b)(4).Collar fixation pin 2pk 40mm, quick release (cat# 201-78-82 / serial# (b)(4).Stryker kms1312.M62 90x13x1.19mm (cat# 203-96-50 / serial# (b)(4).Threaded pin size 2.3 collared 2pk (cat# 521-78-23 / serial# (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information: please disregard this report as it was submitted in error.Event will be reported under report 1038671-2022-01539.
|
| |
|