Model Number 9-ASD-028
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6) 2022, a 28mm amplatzer septal occluder was selected for implantation using a 10f amplatzer torqvue delivery system.During the procedure, during deployment, the occluder deformed into a cobra shape.The occluder was replacer with another 28mm amplatzer septal occluder and the delivery system was replaced with another 10f amplatzer torqvue delivery system.The patient is reported to be discharged.The device was not released inside patient anatomy, there was no angulation or kink noted in the delivery system, and there was no interaction with cardiac structures during deployment.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Use of the incorrect size delivery system, anatomical interference, and angulation or kink in the delivery system upon deployment are potential causes of the reported event.Information from the field indicated that the correct size delivery sheath was used.The patient did not have any anatomical interference and there was no angulation or kink in the delivery system upon deployment.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15834388
Report Number2135147-2022-02130
Device Sequence Number6096716
Product Code MLV
UDI-Device Identifier05415067019325
UDI-Public05415067019325
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number9-ASD-028
Device Catalogue Number9-ASD-028
Device Lot Number6280760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/28/2022
Supplement Date Received by Manufacturer01/31/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date02/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/16/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMPLATZER TORQVUE DELIVERY SYSTEM, 9-ITV10F45/80
Patient SexUnknown
-
-