The customer reported to olympus during a diagnostic egd procedure for epigastric abdominal pain with suspected gerd status post roux-en-y, the subject device failed.The subject device was unable to cut through the suture in the patient and could not be extracted.The customer reported the doctor was able to extract the scope being used and will use another scope to continue the procedure.Customer reported that the procedure was completed and they were able to utilize a needle knife to cut the stitch that the device was clamped on.There were no procedural or anatomical challenges that could have contributed to the reported issue.The patient did not experience any adverse effects due to this occurrence.The patient's current condition is reported as stable, at baseline.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the reported phenomenon occurred because the user attempted to cut sutures that were not intended for use.However, the root cause could not be identified.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿before each case, prepare and inspect the instrument as instructed below.Inspect other equipment to be used with the instrument as instructed in their respective instruction manuals.Should the slightest irregularity be suspected, do not use the instrument; contact olympus.Damage or irregularity may compromise patient or user safety, such as punctures, hemorrhages or mucous membrane damage and may result in more severe equipment damage.¿ ¿do not try to cut the loop that is not positioned on both edges of the loop hanger as plumb as possible for the blade.It may make cutting the loop impossible or result in the loop getting caught in the distal end of the instrument, which could make it difficult or impossible to remove from the patient.In this case, use pliers to cut the insertion portion of the instrument where it extends from the biopsy valve of the endoscope.Remove the endoscope from the body, then reinsert the endoscope and cut the loop with a spare loop cutter.¿ ¿do not use the instrument to cut any objects other than the surplus of olympus loop (e.G., maj-254, maj-340).If anything, other than the surplus of the loop is cut, the blades may be damaged and unable to perform properly, the cut object may be caught in the tip of the instrument, and it may become difficult to safety remove the instrument from the body.Such objects other than the surplus of loop may include, stent wire, sewing threads, and loop stoppers.¿ ¿do not try to cut the loop that is not positioned on both edges of the loop hanger as plumb as possible for the blade.It may make cutting the loop impossible or result in the loop getting caught in the distal end of the instrument, which could make it difficult or impossible to remove from the patient.In this case, use pliers to cut the insertion portion of the instrument where it extends from the biopsy valve of the endoscope.Remove the endoscope from the body, then reinsert the endoscope and cut the loop with a spare loop cutter.¿ ¿do not use the loop cutter to cut anything other than the loop.If you cut any other object, it may get caught in the distal end.This could make it difficult or impossible to remove it from the patient.¿ ¿do not cut the loop unless you have a clear endoscopic field of view.This could cause patient injury, such as punctures, hemorrhages, or mucous membrane damage.It may also damage the endoscope and/or instrument.¿ this supplemental report includes information added to d8.Olympus will continue to monitor field performance for this device.
|