• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LITE; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LITE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 70805-70
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer¿s product has been requested for investigation.A follow-up report will be filed once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 88 mg/dl, 110 mg/dl, 398 mg/dl and 127 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 88 mg/dl, 110 mg/dl, 398 mg/dl and 127 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhr for the freestyle lite meter were reviewed and the dhr showed the freestyle lite meter passed all tests prior to release.Dhr for the freestyle strips were reviewed and the dhr showed the freestyle strips passed all tests prior to release.Retain testing was performed for freestyle strips and all units performed in specification and passed.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Additional Manufacturer Narrative
Meter (b)(6) has been returned and investigated.Visual inspection has been performed on the returned meter and no issues were observed.Meter powered on with button and with strip insertion.Control solution testing was performed and no issues were observed.All results were within range specification.No malfunction or product deficiency was identified.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
Customer reported receiving erratic readings on their adc device.Customer reported receiving readings of 88 mg/dl, 110 mg/dl, 398 mg/dl and 127 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LITE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15834763
Report Number2954323-2022-42599
Device Sequence Number5200369
Product Code NBW
UDI-Device Identifier00699073708274
UDI-Public00699073708274
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2024
Device Model Number70805-70
Device Catalogue Number70827
Device Lot Number1183843
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer12/13/2022
01/03/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/21/2022
01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-