• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTSINE TECHNOLOGIES LTD PACKAGE,360P,PP01,EN,360-STR-US-10; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTSINE TECHNOLOGIES LTD PACKAGE,360P,PP01,EN,360-STR-US-10; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number 360P
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
 
Event or Problem Description
The customer contacted heartsine to report that their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
 
Additional Manufacturer Narrative
Heartsine has made multiple attempts to contact the customer regarding the status of their device, but no response has been received from the customer.The device was not returned to heartsine for investigation.No additional information will be forthcoming at this time.The cause of the reported issue could not be determined.If the device is returned to the manufacturer the investigation will be reopened.H3 other text : device not returned for evaluation.
 
Event or Problem Description
The customer contacted heartsine to report that their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
 
Event or Problem Description
The customer contacted heartsine to report that their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
 
Additional Manufacturer Narrative
Heartsine's investigation of the device confirmed the reported fault as upon return the device was unable to be powered up.The reported fault was attributed to a blown fuse in the returned pedi-paks, which had resulted in an open circuit across the battery packs.The device was scrapped by heartsine and the customer was provided with a replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE,360P,PP01,EN,360-STR-US-10
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
todd bandy
207 airport road west
belfast BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key15835537
Report Number3004123209-2022-00153
Device Sequence Number20190252
Product Code MKJ
UDI-Device Identifier05060167120688
UDI-Public05060167120688
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P160008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number360P
Device Catalogue Number360-STR-US-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer01/16/2023
03/22/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date02/13/2023
04/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-