| Model Number |
360P |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Heartsine has requested return of the device for investigation.Upon completion, the conclusions will be submitted in a follow-up report.
|
| |
|
Event or Problem Description
|
|
The customer contacted heartsine to report that their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Heartsine has made multiple attempts to contact the customer regarding the status of their device, but no response has been received from the customer.The device was not returned to heartsine for investigation.No additional information will be forthcoming at this time.The cause of the reported issue could not be determined.If the device is returned to the manufacturer the investigation will be reopened.H3 other text : device not returned for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer contacted heartsine to report that their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted heartsine to report that their device would not power on.In this state the device may not be able to deliver defibrillation therapy if needed.There was no report of patient use associated to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Heartsine's investigation of the device confirmed the reported fault as upon return the device was unable to be powered up.The reported fault was attributed to a blown fuse in the returned pedi-paks, which had resulted in an open circuit across the battery packs.The device was scrapped by heartsine and the customer was provided with a replacement device.
|
| |
|