• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA HP SYSTEM HANDLE; KNEE INSTRUMENT : HANDLES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA HP SYSTEM HANDLE; KNEE INSTRUMENT : HANDLES Back to Search Results
Model Number 2024-99-111
Medical Device Problem Codes Device-Device Incompatibility (2919); Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the handle broken, won't connect.No pieces broke off just loose.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4), investigation summary : the device associated with this report was returned to depuy synthes for evaluation.Examination of the returned device found the locking mechanism would not open and close as intended, confirming the reported allegation.Visual inspection found nicks near the proximal part of the metal tip of the device.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP SYSTEM HANDLE
Common Device Name
KNEE INSTRUMENT : HANDLES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr
warsaw, IN 46580
3035526892
MDR Report Key15837283
Report Number1818910-2022-23399
Device Sequence Number17666325
Product Code LXH
UDI-Device Identifier10603295084174
UDI-Public10603295084174
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2024-99-111
Device Catalogue Number202499111
Device Lot NumberSO2031454
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age62 YR
Patient SexFemale
-
-