Catalog Number 03.019.008
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, the patient underwent the open reduction internal fixation with multiloc for the proximal humeral fracture.During drilling for distal locking screw, the drill interfered with the nail, and the surgeon could not do drilling.There was no problem before drilling.The connecting screws did not loosen.Drilling for the two distal screw holes was not possible due to interference.Therefore, drilling was performed by checking the misalignment with the hole through the k-wire and manually loading the sleeve in that direction.The surgery was completed successfully within 30 minutes delay.The surgeon guessed that good bone quality may have caused flexure due to loading of the nail at the connection between the anterior margin of the acromion and the device during insertion via ri approach.No further information is available.This report is for one (1) aim-arm lat f/multiloc phn.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: investigation summary product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device history lot part # 03.019.008, lot # l805362, manufacturing site: werk hagendrof, release to warehouse date: 29 may 2018, supplier: n/a, expiration date: n/a.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.B5: event description updated.E3: initial reporter is a physician.
 
Event or Problem Description
There are no pieces left in the patient.The surgeon confirmed this by x-ray.The patient status was reported to be stable.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the aim-arm lat f/multiloc phn.A dimensional inspection was performed for the aim-arm lat f/multiloc phn and met specifications.A functional test could not be performed as the mating device was not returned.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the aim-arm lat f/multiloc phn was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed? yes, reviewed.Dimensional inspection: conforming.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING ARM/LAT/F/MULTILOC PROXIMAL HUMERAL NAIL
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HAGENDROF
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15837624
Report Number8030965-2022-10124
Device Sequence Number20190404
Product Code HSB
UDI-Device Identifier07611819425583
UDI-Public(01)07611819425583
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.019.008
Device Lot NumberL805362
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer11/30/2022
01/03/2023
02/20/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/20/2022
01/26/2023
03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/29/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - GUIDE/COMPRESSION/K-WIRES; UNK - NAILS: MULTILOC HUMERAL; UNK - SCREWS: NAIL DISTAL LOCKING; UNK - SCREWS: NAIL DISTAL LOCKING; UNKNOWN DRILL BITS
Patient Age32 YR
Patient SexMale
-
-