Model Number 1651-84
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was received for investigation and the reported event was confirmed.The balloon was observed to be ruptured.The precise root cause of the ruptured balloon could not be determined.The ifu provides the following warning regarding the possibility of the balloon rupturing during use: arterial rupture or balloon failure due to sharp calcified plaque may occur.The possibility of balloon rupture must be taken into account when considering the risks involved in the craterization procedure.Review of the lot record did not find any issues that would contribute to the event.Additionally, we have not received any other complaints of a similar nature for devices from this lot.There was no injury to the patient as a result of the incident.The surgeon used another device to complete the procedure.
 
Event or Problem Description
A tuftex over-the-wire embolectomy catheter was being used to remove a thrombus.During pre-use check there were no issues.After inserting the catheter into the blood vessel and pulling the catheter for the thrombectomy the balloon ruptured.Another product was used as a substitute to complete the procedure.No injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUFTEX OVER-THE-WIRE EMBOLECTOMY CATHETER
Common Device Name
EMBOLECTOMY CATHETER
Manufacturer (Section D)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer (Section G)
LEMAITRE VASCULAR, INC.
63 second ave
burlington MA 01803
Manufacturer Contact
jeffrey oddy
63 second ave
burlington, MA 01803
7814251682
MDR Report Key15837800
Report Number1220948-2022-00121
Device Sequence Number12119449
Product Code DXE
UDI-Device Identifier00840663100743
UDI-Public(01)00840663100743
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022145
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2026
Device Model Number1651-84
Device Catalogue Number1651-84
Device Lot NumberOTW4602
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2022
Initial Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-