| Catalog Number |
0250040112 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Codes |
Superficial (First Degree) Burn (2685); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient suffered minor burn due to arcing.There was no medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for investigation and the reported failure mode was not confirmed.Alleged failure: arcing causing patient burn probable root cause: heat conducted by device use error insulation damage or inadequate insulation from manufacturing/ servicing lot number is unknown; therefore, manufacture date can not be confirmed.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient suffered minor burn due to arcing.There was no medical intervention.
|
| |
|