• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1080-080
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat the av shunt with diffuse mild calcification, mild tortuosity and 70% stenosis.The 8x80 mm armada 35 percutaneous transluminal angioplasty (pta) catheter ruptured during the first inflation at 12 atmospheres during use.A new armada 35 was used to complete the procedure.There were no reported adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported material rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material rupture.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15839579
Report Number2024168-2022-11818
Device Sequence Number17603843
Product Code LIT
UDI-Device Identifier08717648154881
UDI-Public08717648154881
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model NumberB1080-080
Device Catalogue NumberB1080-080
Device Lot Number20421G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-