| Model Number |
1973-03 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was a pinhole rupture in the balloon.A ranger global dcb otw 6.0 x 200mm, 90cm was selected for use in the endovascular therapy (evt) procedure.The target lesion was within the moderately tortuous superficial femoral artery (sfa) and was reported to be 100% stenosed with moderate calcification.The ranger was advanced to the target lesion and inflated for the first time to 6 atm before bursting.The form of the rupture was reported to be a pinhole.The device was retracted without issue.The ranger was replaced and the procedure was completed.There were no reported adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
E1 - initial reporter state: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was initially reported that there was a pinhole rupture in the balloon.However, on 28nov2022, additional information was received that indicated there was no longer any allegation of device deficiency.The initial report of pinhole rupture had been misreported by the sales representative.Therefore, the event no longer meets complaint criteria.
|
| |
|