Model Number 383592
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd nexiva¿ diffusics¿ closed iv catheter system experienced discoloration.The following information was provided by the initial reporter: discoloration of the tip of the diffusics catheter.
 
Additional Manufacturer Narrative
Our quality engineer inspected the 1 photo submitted for evaluation.The reported issue of discolored was not confirmed upon inspection of the photo.Bd cannot confirm the cause of the failure to our manufacturing process since no sample was returned for evaluation.Dhr for this lot was reviewed and there are no internal rejects related to the reported issue by the customer.According to the quality records all the inspections of the sampling plan met the acceptance criteria.There are a frequential inspections for the tip quality no failures or extraordinary events were recorded on the dhr.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd nexiva¿ diffusics¿ closed iv catheter system experienced discoloration.The following information was provided by the initial reporter: discoloration of the tip of the diffusics catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEXIVA¿ DIFFUSICS¿ CLOSED IV CATHETER SYSTEM
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15840582
Report Number9610847-2022-00439
Device Sequence Number5846030
Product Code FOZ
UDI-Device Identifier30382903835929
UDI-Public(01)30382903835929
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K123734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/31/2025
Device Model Number383592
Device Catalogue Number383592
Device Lot Number2201690
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/28/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date01/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-