The article, "rescue atrial septal defect closure with the new gore® cardioform atrial septal defect occluder" was reviewed.It was reported that on an unknown date a 22mm amplatzer septal occluder was selected for implantation.Prior to the procedure, the patient was diagnosed with a right bundle branch block by electrocardiogram (ekg).A transthoracic echocardiogram (tte) showed that the defect was 14mm.The device was successfully implanted.24 hours after the procedure, the patient developed a complete heart block.They administered methylprednisolone therapy, and then the device was removed using a goose neck catheter.The patient recovered and had normalized rhythm.A few months later, a second attempt at sizing was done, and it was found that the defect was 23-24mm.A 44mm atrial septal defect (asd) closure device was selected and successfully implanted.At the six month follow up, the patient was reported to be stable and normal sinus rhythm.No additional information was provided.The primary author of the article is alessandra pizzuto, md, department of pediatric cardiology and guch unit, heart hospital ¿g.Pasquinucci¿, national research council tuscany region ¿g.Monasterio¿, massa, italy.The correspondence author of the article is giuseppe santoro, md, via camillo sorgente, 31, 84125 ¿ salerno, italy.With the corresponding e-mail: santoropino@tin.It.
|