Model Number 60-19326
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Perforation (2001)
Date of Event 10/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that "too fast threaded drill (second device), for which the surgeon pierced the finger of the patient.Injuries requiring healing care.".
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that "too fast threaded drill (second device), for which the surgeon pierced the finger of the patient.Injuries requiring healing care.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed although the device was returned for evaluation, no evidence for the adverse impact to the patient was provided.The device inspection revealed the following: the visual inspection has shown that the received twist drill bit is in a very used condition, the laser markings are partially faded.The color coding is faded and partially missing.The shaft between the color coding and the cutting flutes is strongly worn.There are chatter marks, caused by many contacts between the shaft and the drill sleeve all over the shaft visible.The cutting edges at the flutes are worn, there are nicks and wear marks all over.Also the cutting edge at the forefront are blunt, the edges rounded and compressed in the outer area.A review of the device history for the reported lot did not indicate any abnormalities.The lot in question was released in june 2006 and we are not aware of any other complaint of this nature for the catalog# 60-25126 or 60-25126s.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The investigation has shown that the drill bit was obviously many times used since it was manufactured.The failure was finally caused by the use of a blunt drill bit, by that higher force than normal was required to drill through the bone and therefore it was not possible to stop the drill after the bone was perforated.In relation to this complaint following statements of the variax instruments instructions for use can be pointed out: "warning: twist drills are for single use only." if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWIST DRILL, DIAM.2.5X94MM, WL 59MM, STRYKER SHAFT END
Common Device Name
BIT, DRILL
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
postfach
freiburg D-791 11
GM   D-79111
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15840844
Report Number0008031020-2022-00638
Device Sequence Number5846038
Product Code HTW
UDI-Device Identifier04546540247360
UDI-Public04546540247360
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number60-19326
Device Catalogue Number60-25126
Device Lot NumberW05
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer02/08/2023
Initial Report FDA Received Date11/21/2022
Supplement Report FDA Received Date03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-