Model Number M0036180840
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The device was received.The reported lot number was not confirmed as the packaging was not returned with the device.During visual/microscopic inspection, the coil delivery wire was found to be kinked/bent.The coil introducer sheath was found to be intact.The main coil was found to be broken/fractured.The main coil was found to be kinked/bent.Main coil stretched functional test ¿ not applicable as the defect was not confirmed during visual inspection.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was not confirmed during the analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.The device was analyzed.The main coil and coil delivery wire were found to be kinked/bent.There were no stretches noted.The main coil was found to be broken/fractured.An assignable cause of not confirmed will be assigned to the as reported event of 'main coil stretched' as the defect was not confirmed during the analysis.An assignable cause of procedural factors will be assigned to the as analyzed events of 'main coil broken/fractured during use' and 'main coil kinked/bent' as the defects appear to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural and/or anatomical factors during use.The assignable cause of handling damage will be assigned to the as analyzed event of 'coil delivery wire kinked/bent' as the defect appears to be associated with handling of the product or portion of the product during the procedure.
 
Event or Problem Description
Analysis of returned device revealed that subject main coil was broken/fractured during use.There was no clinical consequence to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XXL 360 8MM X 40CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15840968
Report Number3008881809-2022-00583
Device Sequence Number6909708
Product Code HCG
UDI-Device Identifier07613327466423
UDI-Public07613327466423
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161429
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/27/2024
Device Model NumberM0036180840
Device Catalogue NumberM0036180840
Device Lot Number23345689
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2022
Initial Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-