MAUDE Adverse Event Report: ASAHI INTECC CO., LTD. GAIA PV; PERIPHERAL GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASAHI INTECC CO., LTD. GAIA PV; PERIPHERAL GUIDE WIRE Back to Search Results
Catalog Number PHW18R302P
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing site: asahi intecc (thailand) co., ltd.The reported gaia pv guide wire was returned for evaluation.Elongation of the outer coil and fracture of the core were confirmed on the returned guide wire.The outer coil was elongated for approximately 51cm from the mid solder that was originally located at 55mm from the tip.The ball tip on the very distal end of the elongated outer coil was found intact.The outer coil remained in one piece.The core wire was fractured at approximately 50mm distal to the mid solder.Twisting caused by accumulated torsion and bending were observed proximal to the fracture end of the core wire.On the distal side of the returned guide wire, a bend was also observed.The core fracture end was located at approximately 5mm from the tip.Distal to the fracture end, twisting and bending as seen on the proximal side of the fracture were confirmed.At approximately 11mm proximal to the tip, the end of the inner coil was exposed.Soldering material was attached at the exposed end of the inner coil, indicating that the inner coil remained in one piece.As fracture of the coil was not confirmed and measurement of the core length was in accordance with the product specification, it was concluded that the entire guide wire was returned.Lot history review revealed no anomaly relating to the reported event.No other similar product experience report was received from this lot.Based on the provided information and investigation outcome, it was presumed that torsion generated with torquing wire manipulation in attempts to cross had most likely contributed to the observed core fracture.The applied stress would localize and therefore exceed the product design limit when the tip was being caught and restricted its movement by the lesion and thus it made the core to fracture.Tensile stress generated with removal then made the outer coil to elongate and the inner coil to detach from the solder jointed onto the core.It was concluded that this event was not attributed to product quality.No capa will be taken.Instructions for use (ifu) states: [warnings]: observe movement of this guide wire in the vessels.Before this guide wire is moved or torqued, the tip movement should be examined and monitored under fluoroscopy.Do not move or torque the guide wire without observing corresponding movement of the tip; otherwise, the guide wire may be damaged and/or trauma may occur.In addition, ensure that the distal tip of this guide wire and its location in the vessel are visible during manipulations of the guide wire.Never push, auger, withdraw, or torque this guide wire that meets resistance.Torquing or pushing this guide wire against resistance may cause damage and/or tip separation of this guide wire or direct damage to a vessel.Resistance may be felt and/or observed under fluoroscopy by noting any buckling of the guide wire.If the prolapse of the guide wire tip is observed, do not allow the tip to remain in a prolapsed position; otherwise damage to the guide wire may occur.Determine the cause of resistance under fluoroscopy and take any necessary remedial action.If resistance is felt due to spasm, bending of the guide wire, or due to trap while operating this guide wire in the blood vessel or removing it, do not torque and/or pull the guide wire itself.Stop the procedure.Determine the cause of resistance under fluoroscopy and take appropriate remedial action.If the guide wire is moved excessively, it may break or become damaged, which may cause blood vessel injury or result in fragments being left inside the vessel.When torquing this guide wire inside the blood vessel, do not torque continuously in the same direction.This may cause the guide wire to become damaged or break apart, causing injury to the blood vessel or leaving fragments inside the vessel.When torquing the guide wire, rotate it clockwise and counterclockwise alternately.Do not exceed two rotations (720 degrees) in the same direction.[malfunction and adverse effect]: breakage of the guide wire.
 
Event or Problem Description
It was reported that an asahi gaia pv guide wire was used for a percutaneous peripheral intervention to treat a moderately tortuous, moderately calcified, 75-90% stenosis in the unspecified limb artery.During the procedure, the wire seemed "trapped." when attempting to remove and release the trapped wire, the wire began to unravel.The wire was able to be removed without any pieces being lost.There was no adverse reaction to the patient.The patient was fine after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAIA PV
Common Device Name
PERIPHERAL GUIDE WIRE
Manufacturer (Section D)
ASAHI INTECC CO., LTD.
3-100 akatsuki-cho
seto, aichi 489-0 071
JA  489-0071
Manufacturer (Section G)
ASAHI INTECC CO., LTD.
3-100 akatsuki-cho
seto, aichi 489-0 071
JA   489-0071
Manufacturer Contact
yukako homma
3-100 akatsuki-cho
seto, aichi 489-0-071
JA   489-0071
MDR Report Key15841180
Report Number3003775027-2022-00154
Device Sequence Number5994263
Product Code DQX
UDI-Device Identifier04547327106917
UDI-Public(01)04547327106917(17)221231(10)200118A461
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K163426
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Catalogue NumberPHW18R302P
Device Lot Number200118A461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Initial Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-