| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Weight, ethnicity: unknown/ not provided.Asku.Date of event: unknown, not provided.Best estimate of date of event is between (b)(6) 2022.Email address: unknown/not provided.Device evaluation: product evaluation was not performed because the product has not been returned.The complaint issue reported could not be verified and no product deficiency could be identified.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no other complaints were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an intraocular lens (iol) was explanted from the patient's right eye due to the patient not seeing well with the symfony even with correction for astigmatism.The explanted lens was replaced with a model zcu.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9.Device available for evaluation? yes.Returned to manufacturer on: dec.8, 2022.Section h3.Device evaluated manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the lens was received cut in half and coated in viscoelastic residue.The lens was cleaned and inspected under magnification and lens damage was identified on the outer edge of the optic.No issues that could cause or contribute to the complaint issue were observed.The complaint issue ¿hm-blurry vision¿ was not confirmed.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|