| Model Number |
MI1000 MED-EL CONCERT PIN |
| Medical Device Problem Codes |
Impedance Problem (2950); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow-up report.
|
| |
|
Event or Problem Description
|
|
The user suffered a fall in (b)(6) 2022.Furthermore, affected channels were seen.Re-implantation is being considered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.Reportedly, the recipient experienced a head trauma, which might have weakened the electrode's fixation.However, to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
The user suffered a fall in (b)(6) 2022.Re-implantation is being considered but no date has been set yet.
|
| |
|
Event or Problem Description
|
|
The user suffered a fall in august 2022.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.Also, an impact to the head might have weakened the fixation of the active electrode and could also be a factor for electrode mobility.The investigation results appear to match the problems mentioned in the recipient report.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
|
| |
|