| Model Number |
956-623 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
11/17/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated based on date of awareness.No date was provided in the complaint.
|
| |
|
Event or Problem Description
|
|
According to the complaint, the nurse was not able to activate the needle shield device on the safepico and was pricked by the needle at the end of the arterial sampling.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of awareness is now provided, to be 2022-nov-16.The safepico in this complaint is not available for investigation.But investigations is currently still ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The radiometer investigation has found that the root cause was traced to manufacturing, where the glue from the barcode label melts after the sterilisation process and can stick to the needle shield device of the safepico.
|
| |
|