MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. BIOMET BONE CMENT R 1X40 JP; BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. BIOMET BONE CMENT R 1X40 JP; BONE CEMENT Back to Search Results
Model Number N/A
Device Problems Improper Chemical Reaction (2952); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign - japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Discarded.
 
Event Description
It was reported that during surgery, the cement hardened within 6-7 minutes.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.The product was not returned.Pictures were not provided.The reported event couldn't be confirmed.A retain sample of batch x02cac1701 has been tested in the laboratory under standardized conditions (23°c; rel.Humid >=40 %).No unusual behavior during mixing, handling or setting.The reported behavior of the cement cannot be reproduced.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.This device is used for treatment.Reported event is not related to medical condition.Review of medical record is not applicable.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET BONE CMENT R 1X40 JP
Type of Device
BONE CEMENT
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15842657
MDR Text Key304975001
Report Number3006946279-2022-00118
Device Sequence Number1
Product Code LOD
UDI-Device Identifier00887868358487
UDI-Public(01)00887868358487(17)231031(10)X02CAC1701
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model NumberN/A
Device Catalogue Number110035372
Device Lot NumberX02CAC1701
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/01/2022
Initial Date FDA Received11/22/2022
Supplement Dates Manufacturer Received12/19/2022
Supplement Dates FDA Received01/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-