• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INBONE STEM, TIBIAL, BASE SIZE, 18 MM, RIGHT & LEFT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INBONE STEM, TIBIAL, BASE SIZE, 18 MM, RIGHT & LEFT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 200009902
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Failure of Implant (1924); Unspecified Infection (1930)
Date of Event 10/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the physician believed that he didn¿t engage the tibial stem to the tibial tray correctly.The tibial tray came away from the stem.It was noted that infection occurred around the stem.The surgeon explanted all the components.The patient is healing in preparation for fusion.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
Correction - d9 / h3 (product available to stryker), h3 (device evaluated by mfg) , h6(method code).The reported event could not be confirmed due to lack of additional information.The device inspection revealed the following: visual examination of the returned devices found scratches and deformations around the tibial stem base.Scratches and blemishes can also be seen on the superior side of the talar dome.Part of the tibial stem base can be seen to be chipped off.The poly was found connected to the tibial tray and could not be detached.Remnants of bone cement and ossification are visible around the tibial tray, tibial stem and talar dome.Infections are often caused by a multitude of factors such as comorbidity of the patient, the trauma, the surgical intervention and the aftercare.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If more information is provided, the case will be reassessed.
 
Event or Problem Description
It was reported that the physician believed that he didn¿t engage the tibial stem to the tibial tray correctly.The tibial tray came away from the stem.It was noted that infection occurred around the stem.The surgeon explanted all the components.The patient is healing in preparation for fusion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INBONE STEM, TIBIAL, BASE SIZE, 18 MM, RIGHT & LEFT
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15843562
Report Number3010667733-2022-00400
Device Sequence Number6940673
Product Code HSN
UDI-Device Identifier00840420123848
UDI-Public00840420123848
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K051023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number200009902
Device Catalogue Number200009902
Device Lot Number1733056
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer02/27/2023
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date03/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-