Model Number 1258T/86
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented in clinic for pacemaker implant procedure.During the implant of the left ventricular lead, it was found that the lead failed to be separated from the stylet.The procedure was completed using an alternate lead on (b)(6) 2022.The patient was stable.
 
Manufacturer Narrative
The reported event for difficult to remove the stylet/guidewire was not confirmed.As received, a complete lead was returned in one piece.Final analysis of the lead found no anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key15844184
MDR Text Key304419687
Report Number2017865-2022-46259
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model Number1258T/86
Device Catalogue Number1258T-86
Device Lot NumberA000104404
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/07/2022
Initial Date FDA Received11/22/2022
Supplement Dates Manufacturer Received01/20/2023
Supplement Dates FDA Received01/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-