Model Number 4674
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2022
Event Type  malfunction  
Manufacturer Narrative
This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
 
Event Description
It was reported that the left ventricular (lv) lead was not implanted successfully due to physician had trouble advancing the wire in the lead.This lead was suspected to have damage proximal to fourth electrode.Further, lead was flushed multiple times.This lead was never in service.No adverse patient effects were reported.
 
Event Description
It was reported that the left ventricular (lv) lead was not implanted successfully due to physician had trouble advancing the wire in the lead.This lead was suspected to have damage proximal to fourth electrode.Further, lead was flushed multiple times.This lead was never in service.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY X4 SPIRAL S
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15844567
MDR Text Key306997643
Report Number2124215-2022-48492
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526524554
UDI-Public00802526524554
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/15/2024
Device Model Number4674
Device Catalogue Number4674
Device Lot Number883536
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/10/2022
Initial Date FDA Received11/22/2022
Supplement Dates Manufacturer Received01/03/2023
Supplement Dates FDA Received01/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
-
-