• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. FORTIFY; FORTIFY Ø12MM CORE, HEIGHT 19-25MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. FORTIFY; FORTIFY Ø12MM CORE, HEIGHT 19-25MM Back to Search Results
Model Number 151.051
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported there was a revision surgery to replace a fortify spacer that had collapsed post operatively.
 
Additional Manufacturer Narrative
The device was returned for evaluation.Initial observation of the micro-ct scanned shows tissue inside of the inserter hole touching the lock.The tissue may have caused the lock to move to the unlocked position; however, an exact cause of the reported issue could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY
Common Device Name
FORTIFY Ø12MM CORE, HEIGHT 19-25MM
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key15844918
Report Number3004142400-2022-00168
Device Sequence Number17637860
Product Code MQP
UDI-Device Identifier00889095042894
UDI-Public00889095042894
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K112756
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number151.051
Device Lot NumberBAA341SP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/25/2022
Initial Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age66 YR
Patient SexMale
Patient RaceWhite
-
-