Model Number CYF-VHA
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the bending rubber of the subject device was perforated.There was no patient or user injury reported due to the event.The subject device was returned to an olympus service center for evaluation.Upon inspection and testing of the returned device, it was observed that foreign material was adhered to the insertion tube.This report is being submitted for the malfunction found during device evaluation (foreign material in insertion tube).Additional details have been requested regarding the reported issue.At this time, no additional information has been provided.
 
Additional Manufacturer Narrative
During inspection and testing of the returned device, the foreign material could not be identified.The reported issue (perforated bending rubber) was confirmed and found to be caused by external factors.It was observed that there was a leakage due to holes in the bending rubber.In addition, service found there was a leakage due to holes in the insertion tube, the insertion tube was wrinkled, angle malfunction, there was rattling at the connection between the bending tube and the tip due to broken bending tube, and the rubber cover of the forceps plug mouthpiece was floating.Scratches were observed on the insertion tube, and the universal cord due to external factors.The investigation is ongoing; therefore, a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was confirmed by the user facility, but the foreign material could not be further identified.It was not possible to determine why the material remained in the device, and it was not possible to further determine if the instructions for use were followed.It was speculated that physical damage to the device could have contributed to the cause.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Common Device Name
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key15845321
Report Number3002808148-2022-04533
Device Sequence Number20558300
Product Code FAJ
UDI-Device Identifier04953170310508
UDI-Public04953170310508
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062049
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-