• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER HEMODIALYTSIS NON IMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM; CATHETER HEMODIALYTSIS NON IMP Back to Search Results
Model Number IPN919778
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported during the procedure, the physician found the guidewire kinked and bent.A new kit was opened to finish the procedure and procedure was successfully completed.No patient harm was reported.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned multiple components from a cvc kit and a lidstock for analysis.The guide wire will be analyzed as part of this investigation.Signs of use in the form of biological material were observed on the guide wire body.Visual analysis revealed that the guide wire contained a small kink/offset coil towards the distal end.Microscopic examination confirmed the damage and revealed that the distal and proximal welds were spherical and intact.The kinks in the guide wire measured 575mm from the proximal weld.The guide wire total length measured 602mm, which is within the specification limits of 596mm-604mm per the guide wire product drawing.The guide wire outer diameter measured.838mm, which is within the specification limits of.838mm-.877mm per the guide wire product drawing.The guide wire was able to pass through the returned arrow raulerson syringe (ars)/introducer needle subassembly.Minor resistance was encountered due to the kinking on the guide wire; however, it was able to pass completely through the ars/needle subassembly.Performed per ifu statement, "advance guidewire into arrow raulerson syringe approximately 10 cm until it passes through syringe valves or into introducer needle".A manual tug test confirmed that the distal and proximal welds were secure and intact.A device history record review was performed, and no relevant findings were identified.The ifu provided with the kit informs the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested." the customer report of a kinked guide wire was confirmed through complaint investigation.Visual analysis revealed one major kink on the guide wire.Despite the damage, the guide wire met all relevant dimensional and functional requirements, and a device history record review was performed with no relevant findings.Based on the customer report and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of nature.
 
Event or Problem Description
It was reported during the procedure, the physician found the guidewire kinked and bent.A new kit was opened to finish the procedure and procedure was successfully completed.No patient harm was reported.The patient's condition is reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 12 FR X 16 CM
Common Device Name
CATHETER HEMODIALYTSIS NON IMP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15845906
Report Number3006425876-2022-01080
Device Sequence Number6996371
Product Code MPB
UDI-Device Identifier10801902191158
UDI-Public10801902191158
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K993933
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/02/2023
Device Model NumberIPN919778
Device Catalogue NumberCS-22122-F
Device Lot Number71F21K1372
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer12/07/2022
Initial Report FDA Received Date11/22/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED; NOT REPORTED
Patient SexUnknown
-
-